Q. How do I calibrate my EDAQ unit?A. Please follow these links to update your unit, or contact ATI technical support for more information.
EDAQ Calibration procedures
Q. I have 2 of the same type of EDAQ in a project file and 1 will not came on-line, what do I do?
A. Right mouse click on the EDAQ that is off-line and go to the properties field.
Click the Assign serial number button for a list of EDAQ's and their serial numbers.
Choose the serial number of the EDAQ that is off-line and hit OK.
Q. How fast can I acquire data from my EDAQs?
A. The EDAQ16T samples at 250ms, and running at full sample speed consumes approximately 2% of the CAN-ICP bus.
The EDAQ16AI has 4 channels (CH 1-4) which sample up to 1ms and 12 channels (CH 5-16) which sample at a maximum of 5ms. An EDAQ16AI running at full sample speed consumes approximately 27% of the CAN-ICP bus.
Q. How does data transfer, from the EDAQ to the Hub, occur?
A. ATI proprietary protocol (ICP) over a CAN physical layer.
Q. Are EDAQAI channels single ended or true differential?
A. Differential
Q. How does the EDAQP measure signals?
A. User selectable frequency, period, pulse-width, duty-cycle.
Q. How much current does the ATI EDAQs units draw?
A. EDAQs are around 150mA each
Q. What are the input impedances of the EDAQ16AI units?
A. Channels 1-16
Attenuation Switch Off 10M Ohm
Attenuation Switch On 56K Ohm